Description: The powerful 5820A Oscilloscope Calibrator is a complete, yet flexible solution for oscilloscope calibration. It calibrates virtually all digital and analog oscilloscopes quickly, easily and affordably. The 5820A is compact, portable and rugged. On the bench, its small footprint leaves you plenty of workspace. Outside the lab, it's easy to carry, without the need for extra accessories like pulse heads. Additional features include an auxiliary feed-through channel that lets you work with external generators up to 2.4 GHz. And there's an external frequency reference input that allows you to enhance the 5820A's frequency accuracy by locking it to an external 10 MHz reference.
The 5820A calibrator provides you with all the functions required to fully verify the performance of any oscilloscope with a bandwidth up to 2.1 GHz, checking amplitude, time, and bandwidth. It performs the following tests:
Amplitude (vertical deflection) - verifies the performance of all the deflection-related circuitry for amplitude accuracy in each range, using dc or square wave signals.
Fast Edge (dynamic response) - verifies the vertical amplifiers, attenuators, analog-to-digital converters in DSOs, and all other vertical deflection
circuitry for transient response. It provides a =< 150 ps edge for scopes up to 2.1 GHz.
Time Marks (horizontal timing) - verify frequency, accuracy and linearity with a +/- 0.33 ppm accuracy, more than enough to calibrate the most
demanding digital oscilloscopes. Pulsed sawtooth, square and 20% pulse wave forms are provided. The 5820A also supports the use of an extended
frequency reference for even higher frequency accuracy.
Leveled Sine Wave (bandwidth verification) - a 50 kHz to 2.1 GHz leveled sine wave source with excellent flatness and amplitude accuracy determines
the 3 dB roll-off point of the oscilloscope and detects other discontinuities and drop outs.
This system has the following specifications:
Voltage DC
Range: 0 to +/-6.6V (50O)
Range: +/-0.25%
------------
Range: 0 to +/-130V (1 MO)
Best 1 Year Specification: +/-0.025%
Voltage AC (Squarewave)
Range: 0 to +/-6.6V (50Ohm)
Best 1 Year Specification: +/-0.25%
------------
Range +/-1 mV to +/-130V p-p (1 MOhm)
Best 1 Year Specification: +/-0.05%
Fast Edge
Range: 4 mV to 2.5V p-p (50Ohm)
Best 1 Year Specificationwith (5820A-GHz): =< 150 ps Rise Time
Leveled Sinewave
Range: 50 kHz to 2.1 GHz
Best 1 Year Specification: 50 kHz to 2.1 GHz; +/-1.5% Flatness
Time Markers
Range: 5s to 500 ps; Spike, square, 20% pulse, sine
Best 1 Year Specification: +/-1 ppm
Wave Generator
Range:
1.8 mV to 55V p-p (1 MOhm)
1.8 mV to 2.5V p-p (50Ohm)
0.01 Hz to 100 kHz
Square, Sine, Triangle
Best 1 Year Specification: +/-3
Pulse Generator
Range: 15 mV to 1.5V amplitude
Best 1 Year Specification: +/-5%
--------
Range: 1 ns to 500 ns width 20 ms to 200 ns period
Best 1 Year Specification: +/-0.33 ppm
DC Voltage Measurement
Range: 0 to +/-6.6V
Best 1 Year Specification: +/-0.05%
Current DC
Range: +/-100 mA to +/-100 mA
Best 1 Year Specification: +/-0.25%
Current AC
Range:
100 mA p-p to 100 mA p-p
10 Hz to 100 kHz
Squarewave
Best 1 Year Specification: +/-0.25%
TV TriggerNTSC, SECAM, PAL, PAL-M
Input Impedance Measurement
Range: 40 Ohm to 1.5 MOhm
Best 1 Year Specification: +/-0.1%
--------
Range: 5 pF to 50 pF
Best 1 Year Specification: +/-5%
Overload Measurement
5V to 9V (dc or ac squarewave), 5 to 60 seconds
|